Outcomes in Pump- and CGM-Baseline Use Subgroups in the International Diabetes Closed-Loop (iDCL) Trial.
BACKGROUND
We investigated the potential benefits of automated insulin delivery (AID) among individuals with type 1 diabetes (T1D) in sub-populations of baseline device use determined by continuous glucose monitor (CGM) use status and insulin delivery via multiple daily injections (MDI) or insulin pump.
MATERIALS AND METHODS
In a six-month randomized, multicenter trial, 168 individuals were assigned to closed-loop control (CLC, Control-IQ, Tandem Diabetes Care), or sensor-augmented pump (SAP) therapy. The trial included a two- to eight-week run-in phase to train participants on study devices. The participants were stratified into four subgroups: insulin pump and CGM (pump+CGM), pump-only, MDI and CGM (MDI+CGM), and MDI users without CGM (MDI-only) users. We compared glycemic outcomes among four subgroups.
RESULTS
At baseline, 61% were pump+CGM users, 18% pump-only users, 10% MDI+CGM users, and 11% MDI-only users. Mean time in range 70-180 mg/dL (TIR) improved from baseline in the four subgroups using CLC: pump+CGM, 62% to 73%; pump-only, 61% to 70%; MDI+CGM, 54% to 68%; and MDI-only, 61% to 69%. The reduction in time below 70 mg/dL from baseline was comparable among the four subgroups. No interaction effect was detected with baseline device use for TIR ( = .67) or time below ( = .77). On the System Usability Questionnaire, scores were high at 26 weeks for all subgroups: pump+CGM: 87.2 ± 12.1, pump-only: 89.4 ± 8.2, MDI+CGM 87.2 ± 9.3, MDI: 78.1 ± 15.
CONCLUSIONS
There was a consistent benefit in patients with T1D when using CLC, regardless of baseline insulin delivery modality or CGM use. These data suggest that this CLC system can be considered across a wide range of patients.