Randomized Controlled Trial of Mobile Closed-Loop Control.

2020
https://researcherprofiles.org/profile/274449375
31937608
Kovatchev B, Anderson SM, Raghinaru D, Kudva YC, Laffel LM, Levy C, Pinsker JE, Wadwa RP, Buckingham B, Doyle FJ, Brown SA, Church MM, Dadlani V, Dassau E, Ekhlaspour L, Forlenza GP, Isganaitis E, Lam DW, Lum J, Beck RW, iDCL Study Group
Abstract

OBJECTIVE

Assess the efficacy of inControl AP, a mobile closed-loop control (CLC) system.

RESEARCH DESIGN AND METHODS

This protocol, NCT02985866, is a 3-month parallel-group, multicenter, randomized unblinded trial designed to compare mobile CLC with sensor-augmented pump (SAP) therapy. Eligibility criteria were type 1 diabetes for at least 1 year, use of insulin pumps for at least 6 months, age ≥14 years, and baseline HbA <10.5% (91 mmol/mol). The study was designed to assess two coprimary outcomes: superiority of CLC over SAP in continuous glucose monitor (CGM)-measured time below 3.9 mmol/L and noninferiority in CGM-measured time above 10 mmol/L.

RESULTS

Between November 2017 and May 2018, 127 participants were randomly assigned 1:1 to CLC ( = 65) versus SAP ( = 62); 125 participants completed the study. CGM time below 3.9 mmol/L was 5.0% at baseline and 2.4% during follow-up in the CLC group vs. 4.7% and 4.0%, respectively, in the SAP group (mean difference -1.7% [95% CI -2.4, -1.0]; < 0.0001 for superiority). CGM time above 10 mmol/L was 40% at baseline and 34% during follow-up in the CLC group vs. 43% and 39%, respectively, in the SAP group (mean difference -3.0% [95% CI -6.1, 0.1]; < 0.0001 for noninferiority). One severe hypoglycemic event occurred in the CLC group, which was unrelated to the study device.

CONCLUSIONS

In meeting its coprimary end points, superiority of CLC over SAP in CGM-measured time below 3.9 mmol/L and noninferiority in CGM-measured time above 10 mmol/L, the study has demonstrated that mobile CLC is feasible and could offer certain usability advantages over embedded systems, provided the connectivity between system components is stable.

Journal Issue
Volume 43 of Issue 3