Successful At-Home Use of the Tandem Control-IQ Artificial Pancreas System in Young Children During a Randomized Controlled Trial.

2019
https://researcherprofiles.org/profile/274449424
30888835
Forlenza GP, Ekhlaspour L, Breton M, Maahs DM, Wadwa RP, DeBoer M, Messer LH, Town M, Pinnata J, Kruse G, Buckingham BA, Cherñavvsky D
Abstract

OBJECTIVE

Hybrid closed-loop (HCL) artificial pancreas (AP) systems are now moving from research settings to widespread clinical use. In this study, the inControl algorithm developed by TypeZero Technologies was embedded to a commercial Tandem t:slim X2 insulin pump, now called Control-IQ, paired with a Dexcom G6 continuous glucose monitor and tested for superiority against sensor augmented pump (SAP) therapy. Both groups were physician-monitored throughout the clinical trial.

RESEARCH DESIGN AND METHODS

In a randomized controlled trial, 24 school-aged children (6-12 years) with type 1 diabetes (T1D) participated in a 3-day home-use trial at two sites: Stanford University and the Barbara Davis Center (50% girls, 9.6 ± 1.9 years of age, 4.5 ± 1.9 years of T1D, baseline hemoglobin A1c 7.35% ± 0.68%). Study subjects were randomized 1:1 at each site to either HCL AP therapy with the Control-IQ system or SAP therapy with remote monitoring.

RESULTS

The primary outcome, time in target range 70-180 mg/dL, using Control-IQ significantly improved (71.0% ± 6.6% vs. 52.8% ± 13.5%; P = 0.001) and mean sensor glucose (153.6 ± 13.5 vs. 180.2 ± 23.1 mg/dL; P = 0.003) without increasing hypoglycemia time <70 mg/dL (1.7% [1.3%-2.1%] vs. 0.9% [0.3%-2.7%]; not significant). The HCL system was active for 94.4% of the study period. Subjects reported that use of the system was associated with less time thinking about diabetes, decreased worry about blood sugars, and decreased burden in managing diabetes.

CONCLUSIONS

The use of the Tandem t:slim X2 with Control-IQ HCL AP system significantly improved time in range and mean glycemic control without increasing hypoglycemia in school-aged children with T1D during remote monitored home use.

Journal Issue
Volume 21 of Issue 4